This designed for safe and efficient high-temp; high-pressure fluid loss testing that meets API specifications. The digital system is controllable and accurate
Contact UsThe filter press consists of a heating jacket, Safe Cell, Type J thermocouple probe, pressurizing assemblies, and two power cables. The test cell can be heated to a maximum temperature of 500°F (260°C) with a 100ml sample or 350 ° F (177°C) with a 130ml sample in thirty minutes. Maximum working pressure is 1800 psig (12,410 kPa).
Safety features include a lock and release mechanism that secures the cell in place and maximizes heat transfer. The exterior of the jacket is safe to touch, less than 130° F. The HT4700 has a digital temperature controller for better accuracy and control. The thermocouple probe allows precise temperature measurements. Logging test data is done using a data port on the heating jacket.
HT4700 Filter Press Available Configurations
Part No. 102195986 —Single Ended Safe Cell, CO2 Pressurizing Assembly, 15 ml Backpressure Assembly
Part No. 102196306 —Double Ended Safe Cell, CO2 Pressurizing Assembly 15 ml Backpressure Assembly
Part No. 102197003 —Double Ended Safe Cell, Dual Nitrogen Manifold, 15 ml Backpressure Assembly
Part No. 102197111 —Single Ended Safe Cell, Dual Nitrogen Manifold, 15 ml Backpressure Assembly
Part No. 101631160 – HT4700 Heating Jacket Assembly
Includes: Type J thermocouple probe, and two power cables (115 and 230 volt)